Purpose
Materials and Methods
Eight gastro-jejunostomy (GJ) models were performed using porcine stomach and small intestine. Specimens were divided into group A where the cartridge was in the gastric lumen with the anvil in the jejunal lumen and vice versa in group B. Enterotomy was closed and gastric and jejunal ends were stapled off. Burst pressure (BP) was measured by infusion of methylene blue saline into the GJ model until leak occurred. Site of leak, BP, and total volume (TV) infused were recorded. Compliance (C) was calculated from the equation ΔTV/ΔBP.
Results
The BP was greater in group A compared to group B (18 mmHg (range 15–25) versus 11 mmHg (range 8–15); p = 0.045) despite similar TV between the groups 60.00 mL (range 55.00–65.00) versus 51.25 mL (range 40.00–60.00); p = 0.11). The compliance did not significantly differ between groups A and B (6.38 mL/mmHg (range 4.34–8.59) versus 5.61 mL/mmHg (range 3.93–7.21); p = 0.48).
Conclusion
BP recorded when the stapler cartridge was introduced in the gastric lumen was higher than when it was introduced from the jejunal lumen. The lack of difference in compliance between groups suggests that the difference in BP was due to the difference in the direction of staples.